Comparison of Different Biomaterials for Glaucoma Drainage Devices
Part 2
Ramesh S. Ayyala, MD, FRCS, FRCOphth;
B. Michelini-Norris, PhD;
Ana Flores, MD;
Edward Haller;
Curtis E. Margo, MD, MPH
Arch Ophthalmol. 2000;118:1081-1084.
Background Inflammation associated with biomaterials may contribute to the failure of glaucoma drainage devices.
Objective To compare the inflammatory reaction associated with the insertion of Krupin silicone, Molteno polypropylene, and Acrosof end plates in the subconjunctival space of rabbits.
Methods Similar-sized glaucoma end plates made of 3 different biomaterials were sutured to the sclera in the superotemporal quadrant of the rabbit eye. Thirty eyes of 15 albino New Zealand rabbits were randomly assigned to the 3 groups. Conjunctival vascular hyperemia was graded in a masked fashion among the 3 groups. At the end of 3 weeks, the enucleated eyes were examined histologically and by scanning electron microscopy.
Results Molteno polypropylene was associated with more inflammation both in clinical observations and based on histological grading. Silicone and Acrosof were associated with less intense inflammation. One polypropylene end plate was extruded on day 21.
Conclusions Polypropylene appears to be more inflammatory than silicone. Flexible biomaterials appear to be less inflammatory than rigid ones.
Clinical Relevance Bleb failure following glaucoma drainage device implantation could be related to the biomaterial-associated inflammation. Choosing a biomaterial with the least inflammatory potential might enhance the success rate of the glaucoma drainage device.
From the Departments of Ophthalmology, Tulane University Medical Center, School of Medicine, New Orleans, La, and Medical Center of Louisiana at New Orleans (Dr Ayyala); Departments of Ophthalmology (Drs Michelini-Norris and Flores) and Pathology (Dr Haller), University of South Florida, Tampa, and Department of Ophthalmology, Watson Clinic, Lakeland, Fla (Dr Margo).
THIS ARTICLE HAS BEEN CITED BY OTHER ARTICLES
MR Imaging Features of Giant Reservoir Formation in the Orbit: An Unusual Complication of Ahmed Glaucoma Valve Implantation
Jeon et al.
Am. J. Neuroradiol. 2007;28:1565-1566.
ABSTRACT
| FULL TEXT
A Newly Designed Glaucoma Drainage Implant Made of Poly(styrene-b-isobutylene-b-styrene): Biocompatibility and Function in Normal Rabbit Eyes
Acosta et al.
Arch Ophthalmol 2006;124:1742-1749.
ABSTRACT
| FULL TEXT
|